KBF 720 (E5.2) - Constant climate chamber

The KBF series was particularly designed for absolutely reliable stability tests and precise maintenance of constant climatic conditions. With its large reserve capacity and many optional features, it is designed to meet future challenges for many years to come.

 

SIGMA 1-14

  • Electronically controlled APT.line™ preheating chamber and refrigerating system assuring temperature accuracy and reproducible results
  • Temperature range 0 °C (32 °F) up to 70 °C (158 °F) - without humitiy
  • Temperature range 10 °C (50 °F) up to 70 °C (158 °F) - with humitiy
  • Humidity range 10 % RH to 80 % RH
  • MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
  • Features:
    - User friendly LCD screen
    - Easy-to-read menu guide
    - Integrated electronic chart recorder
    - Variety of options for the graphic display of process parameters
    - Real time clock
  • Electronically controlled humidification and dehumidification system with capacitive humidity sensor
  • Suitable for stability tests according to ICH guideline Q1A (R2)
  • Inner glass doors with sealing
  • Independent adjustable temperature safety device class 3.1 (DIN 12880), with visual and audible temperature alarm
  • Access port with silicone plug Ø 30 mm (1.18 inch), left side
  • Safety connection kit for water supply and drainage, including water hose, total length 6 m (19.7 ft.)
  • Ethernet interface for GLP/GMP and the FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
  • 2 stainless steel racks included
  • BINDER test certificate

 

 

 

Technical Specifications

Exterior dimensions

 

Width (mm/inch)

1249 / 49.2

Height (inclusive feet) (mm/inch)

1924 / 75.8

Depth (mm/inch)

887 / 34.9

Plus door handle, I-panel, connection (mm/inch)

52 / 2.1

Wall clearance rear (mm/inch)

100 / 3.9

Wall clearance side (mm/inch)

100 / 3.9

Steam space volume (l/cu.ft.)

918 / 32.4

Number of doors

2

Number of inner glass doors

2

Interior dimensions

 

Width (mm/inch)

970 / 38.2

Height (mm/inch)

1250 / 49.2

Depth (mm/inch)

576 / 22.7

Interior volume (l/cu.ft.)

698 / 24.7

Racks (number standard/max.)

2 / 6

Load per rack (kg/lbs.)

45 / 99

Permitted total load (kg/lbs.)

150 / 331

Weight of the unit (empty) (kg/lbs.)

315 / 695

Temperature data (without humidity)

 

Temperature range (°C/°F)

0 - 70 / 32 - 158

Temperature variation at 25 °C (77 °F) (±K)

0.2

Temperature variation at 40 °C (104 °F) (±K)

0.2

Temperature fluctuation (± K)

0.1

Max. heat compensation up to 40 °C (104 °F) (W)

600

Climatic data (with humidity)

 

Temperature range (°C/°F)

10 - 70 / 50 - 158

Temperature variation

 

at 25 °C (77 °F) / 60 % RF (±K)

0.2

at 40 °C (104 °F) / 75 % RF (±K)

0.2

Temperature fluctuation

 

at 25 °C (77 °F) / 60 % RF (±K)

0.1

at 40 °C (104 °F) / 75 % RF (±K)

0.1

Humidity range (% RH)

10 - 80

Humidity fluctuation (± % RH)

 

at 25 °C (77 °F) / 60 % RF

<= 1.5

at 40 °C (104 °F) / 75 % RF

<= 1.5

Recovery time after door were open for 30 sec.

 

at 25 °C (77°F) / 60% RH) (Min.)

2

at 40 °C (104 °F) / 75% RH) (Min.)

6

Electrical data

 

Housing protection acc. to EN 60529

IP 20

Nominal voltage (±10%) 50 / 60 Hz (V)

200-240 1N~

Nominal power (kW)

3.1

Energy consumption at 40 °C (104 °F) / 75 % RF (W) 1)

620

Noise level (dB (A))

53

 

 

 

Contact us on 01939 291200 or by email info@sciquip.co.uk