- Centrifuges
- Centrifuge Tubes
- Chillers
- Climatic Chambers
- Cloning Cylinders
- CO2 Incubators
- Centrifugal Concentrators
- Flasks
- Freeze Dryers
- Furnaces
- Hungate Tubes
- Incubators
- Incubated Shakers
- Liquid Handling
- Microfuges
- Ovens
- Pipettors
- Plant Growth Chambers
- Roll-in Incubators/Rollers
- Shaking Incubators
- Shakers
- Sterilization Chambers
- Stirrers/Mixers
- Tensiometers
- Test Chambers
- Spinner Flasks
- Vacuum Ovens
- Vacuum Concentrators
- Vacuum Pumps
- Viscometers
- Water Baths/Thermostats
MKT 240
(E3)
Environmental simulation chamber for complex temperature profiles in the low temperature range
The MKT test chambers meet all requirements for testing under extreme temperature conditions between -70 °C (-94 °F) and 180 °C (356 °F). The outstanding heating and cooling speeds permit even faster temperature changes and more complex test cycles to be performed./p>
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range from -70 °C to 180 °C (-94 °F to 356 °F)
- MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- Features:
- User-friendly LCD display
- Integrated electronic line recorder
- Different diagram possibilities of process parameters
- Real time clock - Heated viewing window with interior illumination
- Variable adjustable high-performance fan
- Programmable condensation protection for test material
- Adjustable ramp functions via program editor
- Independent adjustable temperature safety device, Class 2 (DIN 12880)
- 230 V power socket on the right-side operating panel
- 4 potential-free relay outputs that can be activated via MCS controller
- RS 422 interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
- Access port, Ø 50 mm (2 inch), left side
- 1 stainless steel rack included
- 4 castors (with 2 brakes)
- BINDER test certificate
Technical Specifications
Exterior dimensions |
|
Width (mm/inch) |
1130 / 44.5 |
Height (inclusive feet/castors) (mm/inch) |
1938 / 76.3 |
Depth, excl. 54 mm (2.2 inch) for door handle (mm/inch) |
946 / 37.2 |
Wall clearance rear (mm/inch) |
100 / 3.9 |
Wall clearance side (mm/inch) |
160 / 6.3 |
Viewing window width (mm/inch) |
508 / 20 |
Viewing window height (mm/inch) |
300 / 11.8 |
Number of doors |
1 |
Interior dimensions |
|
Width (mm/inch) |
735 / 28.9 |
Height (mm/inch) |
700 / 27.6 |
Depth (mm/inch) |
443 / 17.4 |
Interior volume (l/cu.ft.) |
228 / 8.1 |
Racks (number standard/max.) |
1 / 6 |
Load per rack (kg/lbs.) |
30 / 66 |
Permitted total load (kg/lbs.) |
70 / 155 |
Weight (empty) (kg/lbs.) |
380 / 839 |
Temperature data |
|
Temperature range (°C/°F) 1) |
-70 - 180 / -94 - 356 |
Temperature fluctuation (± K) |
0.1 - 1.0 |
Temperature variation (± K) 2) |
0.1 - 0.4 |
Mean heating rate acc. IEC 60068-3-5 (K/min.) |
5.4 |
Mean cooling rate acc. IEC 60068-3-5 (K/min.) |
4.2 |
Heating up time from -70 °C up to 180 °C (Min.) 3) |
45 |
Cooling down time from 180 °C up to -70 °C (Min.) 3) |
95 |
Max. Heat compensation (W) |
2500 |
Electrical data |
|
Housing protection acc. to EN 50529 |
IP 20 |
Nominal voltage (+10%) 50 Hz (V) |
400 3N~ |
Nominal power (kW) |
6.5 |
Noise level (ca. dB(A)) |
64 |
Energy consumption 4) at 20 °C (68 °F) (W) |
1650 |