MKT 240
(E3) Environmental simulation chamber for complex temperature profiles in the low temperature range

FP 240

The MKT test chambers meet all requirements for testing under extreme temperature conditions between -70 °C (-94 °F) and 180 °C (356 °F). The outstanding heating and cooling speeds permit even faster temperature changes and more complex test cycles to be performed./p>

  • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
  • Temperature range from -70 °C to 180 °C (-94 °F to 356 °F)
  • MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
  • Features:
    - User-friendly LCD display
    - Integrated electronic line recorder
    - Different diagram possibilities of process parameters
    - Real time clock
  • Heated viewing window with interior illumination
  • Variable adjustable high-performance fan
  • Programmable condensation protection for test material
  • Adjustable ramp functions via program editor
  • Independent adjustable temperature safety device, Class 2 (DIN 12880)
  • 230 V power socket on the right-side operating panel
  • 4 potential-free relay outputs that can be activated via MCS controller
  • RS 422 interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
  • Access port, Ø 50 mm (2 inch), left side
  • 1 stainless steel rack included
  • 4 castors (with 2 brakes)
  • BINDER test certificate

 

 

 

Technical Specifications

 

Exterior dimensions

 

Width (mm/inch)

1130 / 44.5

Height (inclusive feet/castors) (mm/inch)

1938 / 76.3

Depth, excl. 54 mm (2.2 inch) for door handle (mm/inch)

946 / 37.2

Wall clearance rear (mm/inch)

100 / 3.9

Wall clearance side (mm/inch)

160 / 6.3

Viewing window width (mm/inch)

508 / 20

Viewing window height (mm/inch)

300 / 11.8

Number of doors

1

Interior dimensions

 

Width (mm/inch)

735 / 28.9

Height (mm/inch)

700 / 27.6

Depth (mm/inch)

443 / 17.4

Interior volume (l/cu.ft.)

228 / 8.1

Racks (number standard/max.)

1 / 6

Load per rack (kg/lbs.)

30 / 66

Permitted total load (kg/lbs.)

70 / 155

Weight (empty) (kg/lbs.)

380 / 839

Temperature data

 

Temperature range (°C/°F) 1)

-70 - 180 / -94 - 356

Temperature fluctuation (± K)

0.1 - 1.0

Temperature variation (± K) 2)

0.1 - 0.4

Mean heating rate acc. IEC 60068-3-5 (K/min.)

5.4

Mean cooling rate acc. IEC 60068-3-5 (K/min.)

4.2

Heating up time from -70 °C up to 180 °C (Min.) 3)

45

Cooling down time from 180 °C up to -70 °C (Min.) 3)

95

Max. Heat compensation (W)

2500

Electrical data

 

Housing protection acc. to EN 50529

IP 20

Nominal voltage (+10%) 50 Hz (V)

400 3N~

Nominal power (kW)

6.5

Noise level (ca. dB(A))

64

Energy consumption 4) at 20 °C (68 °F) (W)

1650

 

 

 

Contact us on 01939 291200 or by email info@sciquip.co.uk