MK 53 (E2.1)
Environmental test chamber for complex temperature profiles

FP 240

This series covers the classic temperature range between -40 °C (-40 °F) and 180 °C (356 °F) for heat and refrigeration tests – with the added benefit of natural simulation by means of preheating chamber technology and the Horizontal Air Flow Design. Unique technology, developed by BINDER.With these features, the MK series thus meets the highest precision and performance requirements for cyclic temperature tests and presents an intelligent alternative to expensive individual solutions.

  • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
  • Temperature range -40 °C (-40 °F) up to 180 °C (356 °F) at ambient temperature 25 °C (77 °F)
  • MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments o User friendly LCD screen o Easy-to-read menu guide o Integrated electronic chart recorder o Variety of options for the graphic display of process parameters o Real time clock
  • Programmable condensation protection for test material
  • Powerful adjustable fan
  • Access port, Ø 80 mm (3.1 inch), top
  • Heated viewing window with interior lighting
  • Environmental friendly refrigerant R 404a
  • Temperature safety device class 2 (DIN 12880) with visual and audible temperature alarm
  • Printer- and communication interface RS 422 interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
  • 1 stainless steel shelf included
  • BINDER test certificate

 

 

 

Technical Specifications

 

Exterior dimensions

 

Width (mm/inch)

740 / 29.1

Height (incl. feet/castors) (mm/inch)

1242 / 48.9

Depth, excl. 55 mm (2.2 inch) for door handle (mm/inch)

794 / 31.3

Wall clearance rear (mm/inch)

100 / 3.9

Wall clearance side (mm/inch)

160 / 6.3

Viewing window width (mm/inch)

280 / 11.0

Viewing window height (mm/inch)

280 / 11.0

Number of doors

1

Interior dimensions

 

Width (mm/inch)

402 / 15.8

Height (mm/inch)

402 / 15.8

Depth (mm/inch)

330 / 13.0

Interior volume (l/cu.ft.)

53 / 1.9

Racks (number standard/max.)

1 / 5

Load per rack (kg/lbs.)

15 / 33

Permitted total load (kg/lbs.)

40 / 88

Weight (empty) (kg/lbs.)

150 / 331

Temperature data

 

Temperature range (°C/°F)

-40 - 180 / -40 - 356

Temperature variation

 

-40 °C (-40 °F) (± °C)

0.8

-10 °C (14 °F) (± °C)

0.7

0 °C (32 °F) (± °C)

0.4

20 °C (68 °F) (± °C)

0.8

70 °C (158 °F) (± °C)

1.2

150 °C (302 °F) (± °C)

2

Temperature fluctuation (± °C) 2)

0.3

Recovery time after 30 sec door open

 

at -10 °C (14 °F) (Min.)

5

at 70 °C (158 °F) (Min.)

1

at 150 °C (302 °F) (Min.)

5

Mean heating-up rate acc. to factory standard -40°C to 180°C (K/min.)

4.6

Mean cooling rate acc. factory standard 180°C to -40°C (K/min.)

4.1

Heating up time from -40 °C up to 180 °C (Min.)***

47

Cooling down time from 180 °C up to -40 °C (Min.)***

93

Heat Compensation, max. (W)

500

Electrical data

 

Housing protection acc. to EN 50529

IP 20

Nominal voltage (±10 %) 50 Hz (V)

230 V 1N~

Nominal power (kW)

2.6

Energy consumption1) at 20 °C (kW)

1.02

Noise level (ca. dB(A))

59

 

 

 

Contact us on 01939 291200 or by email info@sciquip.co.uk