M 53 Temperature test chamber for complex temperature profiles

M53

Because of its individual programming options and ability to operate at maximum temperatures up to 300 °C (572 °F), the M series is ideally suited for materials testing and aging tests. The heavy-duty air turbine and a programmable exhaust ventilation flap provide rapid heating-up and ensure that the test temperature is maintained absolutely precise at all levels, with minimal spatial fluctuations; performance as never before.

  • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
  • Temperature range 5 °C (9 °F) above ambient temperature up to 300 °C (572 °F)
  • MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
  • Features:
    - User friendly LCD screen
    - Easy-to-read menu guide
    - Integrated electronic chart recorder
    - Variety of options for the graphic display of process parameters
    - Real time clock
  • Adjustable ramp function via program editor
  • Program-controlled ventilation flap
  • High air-exchange rate through high-performance fan
  • Adjustable fan speed
  • Exhaust duct Ø 50 mm (1.97 inch)
  • Temperature safety device class 2 (DIN 12880) with visual alarm
  • Printer and communication interface RS 422 for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
  • Units up to 115 liters are stackable
  • 2 chrome-plated racks included
  • BINDER test certificate

 

 

 

Technical Specifications

 

Exterior dimensions

 

Width (mm/inch)

634 / 25.0

Height (inclusive feet) (mm/inch)

779 / 30.7

Depth (mm/inch)

575 / 22.6

plus door handle and exhaust duct (mm/inch)

90 / 3.5

Wall clearance rear (mm/inch)

100 / 3.9

Wall clearance side (mm/inch)

160 / 6.3

Exhaust duct outer- Ø (mm/inch)

52 / 2.1

Steam space volume (l/cu.ft.)

77 / 2.7

Number of doors

1

Interior dimensions

 

Width (mm/inch)

400 / 15.8

Height (mm/inch)

400 / 15.8

Depth (mm/inch)

330 / 13.0

Interior volume (l/cu.ft.)

53 / 1.9

Shelves, chrome - plated (number standard/max.)

2 / 5

Load per rack (kg/lbs.)

15 / 33

Permitted total load (kg/lbs.)

40 / 88

Weight of the unit (empty) (kg/lbs.)

61 / 135

Temperature data

 

Temperature range, 5 °C (41 °F) above ambient up to (°C / °F)

300 / 572

Temperature variation 1)

 

at 70 °C (± °C)

0.5

at 150 °C (± °C)

1.3

at 300 °C (± °C)

2.8

Temperature fluctuation (± °C)

0.3

Heating-up time 2)

 

to 70 °C (Min.)

5

to 150 °C (Min.)

15

to 250 °C (Min.)

35

Recov. time after door was opened for 30 sec. 2)

 

at 70 °C (Min.)

1

at 150 °C (Min.)

3

at 300 °C (Min.)

5

Electrical data

 

Housing protection acc. to EN 60529

IP 20

Nominal voltage (±10 %) 50 / 60 Hz (V)

230

Nominal power (kW)

1.2

Energy consumption

 

at 70 °C (W)

145

at 150 °C (W)

300

at 300 °C (W)

720

 

 

 

Contact us on 01939 291200 or by email info@sciquip.co.uk