M 53 Temperature test chamber for complex temperature profiles
Because of its individual programming options and ability to operate at maximum temperatures up to 300 °C (572 °F), the M series is ideally suited for materials testing and aging tests. The heavy-duty air turbine and a programmable exhaust ventilation flap provide rapid heating-up and ensure that the test temperature is maintained absolutely precise at all levels, with minimal spatial fluctuations; performance as never before.
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range 5 °C (9 °F) above ambient temperature up to 300 °C (572 °F)
- MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- Features:
- User friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real time clock - Adjustable ramp function via program editor
- Program-controlled ventilation flap
- High air-exchange rate through high-performance fan
- Adjustable fan speed
- Exhaust duct Ø 50 mm (1.97 inch)
- Temperature safety device class 2 (DIN 12880) with visual alarm
- Printer and communication interface RS 422 for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
- Units up to 115 liters are stackable
- 2 chrome-plated racks included
- BINDER test certificate
Technical Specifications
Exterior dimensions |
|
Width (mm/inch) |
634 / 25.0 |
Height (inclusive feet) (mm/inch) |
779 / 30.7 |
Depth (mm/inch) |
575 / 22.6 |
plus door handle and exhaust duct (mm/inch) |
90 / 3.5 |
Wall clearance rear (mm/inch) |
100 / 3.9 |
Wall clearance side (mm/inch) |
160 / 6.3 |
Exhaust duct outer- Ø (mm/inch) |
52 / 2.1 |
Steam space volume (l/cu.ft.) |
77 / 2.7 |
Number of doors |
1 |
Interior dimensions |
|
Width (mm/inch) |
400 / 15.8 |
Height (mm/inch) |
400 / 15.8 |
Depth (mm/inch) |
330 / 13.0 |
Interior volume (l/cu.ft.) |
53 / 1.9 |
Shelves, chrome - plated (number standard/max.) |
2 / 5 |
Load per rack (kg/lbs.) |
15 / 33 |
Permitted total load (kg/lbs.) |
40 / 88 |
Weight of the unit (empty) (kg/lbs.) |
61 / 135 |
Temperature data |
|
Temperature range, 5 °C (41 °F) above ambient up to (°C / °F) |
300 / 572 |
Temperature variation 1) |
|
at 70 °C (± °C) |
0.5 |
at 150 °C (± °C) |
1.3 |
at 300 °C (± °C) |
2.8 |
Temperature fluctuation (± °C) |
0.3 |
Heating-up time 2) |
|
to 70 °C (Min.) |
5 |
to 150 °C (Min.) |
15 |
to 250 °C (Min.) |
35 |
Recov. time after door was opened for 30 sec. 2) |
|
at 70 °C (Min.) |
1 |
at 150 °C (Min.) |
3 |
at 300 °C (Min.) |
5 |
Electrical data |
|
Housing protection acc. to EN 60529 |
IP 20 |
Nominal voltage (±10 %) 50 / 60 Hz (V) |
230 |
Nominal power (kW) |
1.2 |
Energy consumption |
|
at 70 °C (W) |
145 |
at 150 °C (W) |
300 |
at 300 °C (W) |
720 |