Binder VDL115 Safety vacuum drying oven for flammable solvents
A BINDER VDL series chamber ensures maximum safety when drying organic solutents standard with TÜV/GS. The inner chamber is designed according to ATEX guidelines for Zone II 3 G.
- Electronically controlled APT.line™ preheating chamber with 2 expansion racks assuring temperature accuracy and reproducible results
- Temperature range from 15 °C above ambient temperature to 200 °C
- MP controller with 2 programs with 10 sections each or switchable to 1 program with 20 sections
- Integrated weekly program timer with real-time function o Digital temperature setting with an accuracy of one degree o Elapsed time indicator
- Spring-mounted safety glass panel with shatter protection
- Precision-adjustable ventilation valve
- Precision-adjustable inert gas valve with Cross-Flow-Technology
- Safety concept:
- Pressure control device for heating activated < 125 mbar o Over pressure capsuled instrument panel with compressed air connection and maintenance unit o Flame protection gasket
- Analog pressure gauge (displays pressure difference between the inner chamber and ambient pressure)
- Electro polished inner chamber, suction and ventilation tubes, pressure container, expansion racks, and ball valve are made of stainless steel
- Door gasket made of tempered silicone
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
- Measuring port DN 16 in rear panel
- Printer and communication interface RS 422 for APT-COM™ DataControlSystem communication software
- 2 patented, flexible aluminum expansion racks
- All electrical components are decoupled from the inner chamber
- Also available as complete system with module and vacuum pump. Features:
- Reduced noise level o Practical working height o Coordinated system
- BINDER test certificate
Technical Specifications
Exterior dimensions |
|
Width (mm) |
740 |
Height (incl. feet) (mm) |
900 |
Height with optional vacuum module (mm) |
622 |
Height with optional vacuum module (mm) |
1522 |
Depth (mm) |
670 |
Plus door handle, connection (mm) |
100 |
Wall clearance, rear (mm) |
100 |
Wall clearance, side (mm) |
135 |
Interior dimensions |
|
Width (mm) |
506 |
Height (mm) |
506 |
Depth (mm) |
450 |
Interior volume (l) |
115 |
Expansion racks (aluminum) (number standard/max.) |
2/6 |
Distance between the racks (mm) |
68 |
Usable space per rack (width x depth) (mm) |
455 x 440 |
Load per shelf (kg) |
20 |
Permitted total load (kg) |
65 |
Weight (empty) (kg) |
153 |
Temperature range |
|
Temperature range approx. 15 °C above ambient temperature to (°C) |
200 |
Temperature variation 1) |
|
100 °C (± K) |
3,5 |
200 °C (± K) |
9 |
Temperature fluctuation (± K) 1) |
0,1 |
Heating up time 1), 2) switch position II |
|
to 100 °C (min.) |
155 |
to 200 °C (min.) |
200 |
Vacuum connection with small flange (DN mm) |
16 |
Measuring access port with small flange (DN mm) |
16 |
Inert gas connection with flow limiter thread (RP) |
3/8 |
Air pressure connection for pressurized apparatus (tube clip Ø / mm) |
8 |
Permitted end vacuum (mbar) |
0,01 |
Leak rate (max. bar 1/h) |
0,01 |
Electrical data |
|
IP protection class acc. to EN 60529 |
IP 54 |
Voltage (± 10%) 50 / 60 Hz |
230 |
Nominal power (kW) |
1,7 |
Energy consumption |
|
at 100 °C (W) |
250 |
at 200 °C (W) |
785 |
1) Value with aluminum racks
2) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. These average values have been determined according to the BINDER factory standard, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. Differing ambient temperatures and production-related device-specific variances can lead to varying technical data.
Therefore, we recommend individual customer-specific calibration or validation for applications on the limit of the permitted ambient temperature range.
Contact us on 01939 291200 or by email info@sciquip.co.uk