Binder VDL115 Safety vacuum drying oven for flammable solvents

A BINDER VDL series chamber ensures maximum safety when drying organic solutents standard with TÜV/GS. The inner chamber is designed according to ATEX guidelines for Zone II 3 G.

  • Electronically controlled APT.line™ preheating chamber with 2 expansion racks assuring temperature accuracy and reproducible results
  • Temperature range from 15 °C above ambient temperature to 200 °C
  • MP controller with 2 programs with 10 sections each or switchable to 1 program with 20 sections
  • Integrated weekly program timer with real-time function o Digital temperature setting with an accuracy of one degree o Elapsed time indicator
  • Spring-mounted safety glass panel with shatter protection
  • Precision-adjustable ventilation valve
  • Precision-adjustable inert gas valve with Cross-Flow-Technology
  • Safety concept:
  • Pressure control device for heating activated < 125 mbar o Over pressure capsuled instrument panel with compressed air connection and maintenance unit o Flame protection gasket
  • Analog pressure gauge (displays pressure difference between the inner chamber and ambient pressure)
  • Electro polished inner chamber, suction and ventilation tubes, pressure container, expansion racks, and ball valve are made of stainless steel
  • Door gasket made of tempered silicone
  • Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
  • Measuring port DN 16 in rear panel
  • Printer and communication interface RS 422 for APT-COM™ DataControlSystem communication software
  • 2 patented, flexible aluminum expansion racks
  • All electrical components are decoupled from the inner chamber
  • Also available as complete system with module and vacuum pump. Features:
  • Reduced noise level o Practical working height o Coordinated system
  • BINDER test certificate

 

 

 

Technical Specifications

 

Exterior dimensions

Width (mm)

740

Height (incl. feet) (mm)

900

Height with optional vacuum module (mm)

622

Height with optional vacuum module (mm)

1522

Depth (mm)

670

Plus door handle, connection (mm)

100

Wall clearance, rear (mm)

100

Wall clearance, side (mm)

135

Interior dimensions

Width (mm)

506

Height (mm)

506

Depth (mm)

450

Interior volume (l)

115

Expansion racks (aluminum) (number standard/max.)

2/6

Distance between the racks (mm)

68

Usable space per rack (width x depth) (mm)

455 x 440

Load per shelf (kg)

20

Permitted total load (kg)

65

Weight (empty) (kg)

153

Temperature range

Temperature range approx. 15 °C above ambient temperature to (°C) 

200

Temperature variation 1)

 

100 °C (± K)

3,5

200 °C (± K)

9

Temperature fluctuation (± K) 1)

0,1

Heating up time 1), 2) switch position II

 

to 100 °C (min.)

155

to 200 °C (min.)

200

Vacuum connection with small flange (DN mm)

16

Measuring access port with small flange (DN mm)

16

Inert gas connection with flow limiter thread (RP)

3/8

Air pressure connection for pressurized apparatus (tube clip Ø / mm)

8

Permitted end vacuum (mbar)

0,01

Leak rate (max. bar 1/h)

0,01

Electrical data

IP protection class acc. to EN 60529

IP 54

Voltage (± 10%) 50 / 60 Hz

230 

Nominal power (kW)

1,7

Energy consumption

 

at 100 °C (W)

250

at 200 °C (W)

785

 

1) Value with aluminum racks 2) To 98% of the set value

All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. These average values have been determined according to the BINDER factory standard, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. Differing ambient temperatures and production-related device-specific variances can lead to varying technical data. Therefore, we recommend individual customer-specific calibration or validation for applications on the limit of the permitted ambient temperature range.

 

 

 

Contact us on 01939 291200 or by email info@sciquip.co.uk