INCUCELL/INCUCELL V with natural forced air convection

Suitable for safe treatment of microbiological cultures. The INCUCELL line produces no noise and provides a very soft air convection within the chamber , the variant INCUCELL V (with a fan) has an advantage of more precise temperature distribution with small deviations. These devices can be used especially in biological and microbiological laboratories, quality tests in pharmacy, cosmetics and testing in veterinary medicine and food processing industry.
- Volume: 22, 55, 111, 222, 404, 707 litres
- Working temperature:
INCUCELL:
5 °C above ambient temperature up to 70,0 °C / 99,9 °C - INCUCELL V:
10 °C above ambient temperature up to 70,0 °C/ 99,9 °C - Inner glass door
Interior:
stainless steel, mat. No. 1.4301 (AISI 304)
The versatile standard line with microprocessor control unit

- 3 programs
- RS 232 – interface for printer or PC-communication
- delayed heating start and stop function
- acoustic and visual alarm
- time range 99 hours 59 minutes
- digital safety thermostat class 3
- manual control of the air flap
(the suction and air exhaust only INCUCELL V, the air exhaust only INCUCELL)
Options
- access ports ? 25, 50, 100 mm (?100 mm is not available for 22-litres volume)
- door lock
- left door versions (excluded volume 22 and 707 litres)
- measurement of material´s temperature with an independent movable sensor PT 100 (with indication on a display or PC)
- communication software WarmComm
- HEPA-filter for installation in an air inlet (only INCUCELL V )
- expansion of working temperature range up to 99.9 °C
- arragement against drying-up of culture mediums and tissue cultures (INCUCELL V only)
- stainless steel casing of the device
The high-tech comfort line with multi-functional microprocessor control unit

- 6 programs
- chip card system for individual program storage
- RS 232- interface for printer or PC-communication
- delayed heating start and stop function
- acoustic and visual alarm
- time range 0–16 years with 1 min-intervals
- digital safety thermostat class 3
- real time
- programming temperature ramps
- heating sequences
- programme cycles
- adjustable ventilation rate 10 to 100 % (only INCUCELL V)
- manual control of the air flap (the suction and air exhaust only INCUCELL V, the air exhaust only INCUCELL)
Options
- access ports ? 25, 50, 100 mm (? 100 mm is not available for 22-litres volume)
- door lock
- left door versions (excluded volume 22 and 707 litres)
- HEPA-filter for installation in air inlet (only INCUCELL V )
- timer programmable water protected inner socket (only INCUCELL V)
- communication software WarmComm
- potential-free alarm contact
- measurement of material´s temperature with an independent movable sensor PT 100 (with indication on a display or PC)
- expansion of working temperature range up to 99.9 °C
- arragement against drying-up of culture mediums and tissue cultures
- high temperature decontamination (excluded volume 404 and 707 litres)
- stainless steel casing of the device
Specifications
|
Model |
22 |
55 |
111 |
222 |
404 |
707 |
|
Interior of stainless steel material No. 1.4301 |
volume |
cca ltrs |
22 |
55 |
111 |
222 |
404 |
707 |
width |
cca mm |
240 |
400 |
540 |
540 |
540 |
940 |
|
depth |
cca mm |
320 |
370 |
370 |
520 |
520 |
520 |
|
height |
cca mm |
295 |
350 |
530 |
760 |
1410 |
1410 |
|
Trays
|
racks |
max. No. |
4 |
4 |
7 |
10 |
19 |
19 |
| standard equipment |
pcs. included |
2 |
2 |
2 |
2 |
2 |
2 |
|
distance between guide rails |
mm |
60 |
70 |
70 |
70 |
70 |
70 |
|
storage area (w x d) |
mm |
185x265 |
380x335 |
520x335 |
520x485 |
520x485 |
920x485 |
|
Maximal weight of the load*)
|
per tray |
max. kg |
10 |
20 |
20 |
30 |
30 |
50 |
inside the oven |
max. kg |
25 |
50 |
50 |
70 |
100 |
130 |
|
Door |
|
No. |
1 |
1 |
1 |
1 |
1 |
2 |
External dimensions (including door and handle)
|
width |
cca mm |
406 |
620 |
760 |
760 |
760 |
1160 |
depth |
cca mm |
580 |
640 |
640 |
790 |
790 |
790 |
|
height (incl. Foots and Rolls) |
cca mm |
604F |
680F |
860F |
1090F |
1910R |
1910R |
|
diameter of the air branch outer/inner |
cca mm |
52/49 |
52/49 |
52/49 |
52/49 |
52/49 |
52/49 |
|
Package dimensions (three layers carton)
|
width depth height (incl. palette) |
cca mm |
465 |
700 |
830 |
830 |
830 |
1230 |
depth |
cca mm |
665 |
730 |
730 |
730 |
860 |
860 |
|
height (incl. palette) |
cca mm |
655 |
880 |
1050 |
1280 |
2070 |
2080 |
|
Weight
|
netto |
cca kg |
31 |
55 |
75 |
100 |
150 |
215 |
brutto |
cca kg |
36 |
66 |
87 |
116 |
175 |
240 |
|
Working temperature (beginning of the regulation) **) |
IC/IC V |
from 5/10 °C above ambient temp. to °C |
70/99,9 |
70/99,9 |
70/99,9 |
70/99,9 |
70/99,9 |
70/99,9 |
Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door |
space deviation IC/IC V |
cca ± °C |
1,1/0,4 |
< 0,5/ |
< 0,5/ |
< 1/< 0,3 |
< 1/< 0,8 |
< 1,5/< 1,5 |
time variation |
cca ± °C |
0,4/0,1 |
< 0,2 |
< 0,2 |
< 0,2 |
< 0,2 |
< 0,2 |
|
Time required to reach 37 °C with closed air flap and voltage 230 V |
IC/IC V |
cca min. |
70/8 |
49/41 |
57/51 |
79/66 |
41/38 |
59/51 |
Heat emission at |
|
cca W |
20/70 |
30 |
45 |
45 |
65 |
85 |
Air exchange speed at 37 °C |
IC/IC V |
cca/hour |
5/45 |
5/45 |
5/49 |
5/24 |
5/18 |
5/12 |
Electricity -mains 50/60 Hz
|
max. power input IC/IC V |
kW |
0,24/0,96 |
0,3/0,7 |
0,3/0,7 |
0,5/0,7 |
0,9/1,3 |
0,9/1,3 |
stand by |
W |
5 |
5 |
5 |
5 |
5 |
5 |
|
current IC/IC V |
A |
2,2/8,4 1,1/4,2 |
2,6/6 1,3/3 |
2,6/6 1,3/3 |
4,4/6 2,2/3 |
7,8/11,3 3,9/5,7 |
7,8/11,3 |
|
nominal voltage |
V |
115 |
115 |
115 |
115 |
115 |
115 |
|
*) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.
**) Standard type is up to 70 °C, optional type is up to 99,9 °C.
Note: All technical data are related to 22 °C ambient temperature and ±10 % voltage swing (if not specified).