VACUCELL with vacuum

Temperature sensitive, easy decomposable or oxidative materials can be dried very tenderly in VACUCELL vacuum drying ovens, where there is the opportunity of extrusion of air by inert gas. Also complicated components with hardly accessible hollow spaces are drying quick and affectively in VACUCELL ovens.

  • Volume:
    22, 55, 111 litres
  • Working temperature:
    5°C above ambient temperature up to 200 °C
  • Door window Intergrated duct for sensors etc. (? 40 mm) Inert gas connection Needle valve for fine dosing Pressure resistant inner chamber Safety valve-door VENTIFLEX Interior:
    stainless steel, mat. No.1.4571 (AISI 316 Ti)

 

 

 

The versatile standard line with microprocessor control unit
  • 3 programs
  • RS 232 – interface for printer or PC-communication
  • delayed heating start and stop function
  • acoustic and visual alarm
  • time range 99 hours 59 minutes
  • digital safety thermostat class 2

 

 

Options
  • comunication software WarmComm
  • base box Vacustation
  • chemical resistant vacuum pump
  • chemical resistang vacuum pump with inlet separator and exhaust condenser
  • external vacuum control system
  • stainless steel casing of the device
  • measurement of material´s temperature with an independent movable sensor PT 100 (with indication on a display or PC)
  • electronic measurement of pressure with indication on a display

 

The high-tech comfort line with multi-functional microprocessor control unit
  • 6 programs
  • chip card system for individual program storage
  • RS 232- interface for printer or PC-communication
  • delayed heating start and stop function
  • acoustic and visual alarm
  • time range 0–16 years with 1 min-intervals
  • digital safety thermostat class 2
  • real time
  • programming temperature ramps
  • heating sequences
  • programme cycles

 

Options
  • base box Vacustation
  • chemical resistant vacuum pump
  • chemical resistang vacuum pump with inlet separator and exhaust condenser
  • built-in vacuum control system
  • communication software WarmComm
  • potential-free alarm contact
  • measurement of material´s temperature with an independent movable sensor PT 100 (with indication on a display or PC)
  • stainless steel casing of the device
  • electronic measurement of pressure with indication on a display

 

Specifications

 

 

Model

22

55

111

Interior of stainless steel material DIN 1.4571 (AISI316TI)

 

 

 

volume

cca litres

22

55

111

width

cca mm

340

400

540

depths

cca mm

260

320

410

height

cca mm

300

430

480

Shelves

 

Shelves guides-impressions in the chamber side walls

number max.

5

8

9

number

Pieces suppl

2

2

2

Shelves distance

height

mm

40

40

40

Usefull dim. of shelf

width x depths

mm

280 x 236

340 x 296

480 x 386

Max. permissible load of the shelves

 

one shelf

kg

20

25

25

totally per unit

kg

35

45

65

External dimensions (including door and handle)

 

 

width

cca mm

560

620

760

depths

cca mm

490

550

640

height

cca mm

700

830

880

Package dimensions (three layers carton)

 

 

width

cca mm

740

830

830

depths

cca mm

615

635

730

height (incl. palette)

cca mm

915

1010

1070

Weight

 

netto

cca kg

65

98

130

brutto

cca kg

76

110,5

144,5

Electric parameters – mains 50/60 Hz

 

 

 

max. input

kW

0,8

1,2

1,8

input in stand by mode

W

5

5

5

current

A

3,5

5,2

7,8

nominal voltage

V

230

230

230

Temperature data Working temp (regulation start)

from 5 °C over ambient temp to °C

 

200

200

200

Temp. deviations acc. to DIN 12 880 Part 2 from working temp – Al shelves pressure 5–10 mbar** 

 

 

at 100 °C

accuracy in space

do ± °C

2

2

3

at 200 °C

accuracy in space

do ± °C

<5

<6

<7

accuracy in time

do ± °C

0,4

0,4

0,4

Temp. deviations acc. to DIN 12 880 Part 2 from working temp – ss shelves pressure 5–10 mbar** 

 

 

at 100 °C

accuracy in time

do ± °C

10

10

11

at 200 °C

accuracy in time

do ± °C

18

23

*

accuracy in time

do ± °C

0,5

1,0

1,0

Time of rise onto 98 % voltage 230 V – (Al shelves pressure 5–10 mbar)

 

onto temp 100 °C

min

60

65

110

onto temp 200 °C

min

80

85

130

Time of rise onto 98 % voltage 230 V – steel shelves, pressure 5–10 mbar

 

onto temp 100 °C

min

130

140

170

onto temp 200 °C

min

170

180

220

Heat radiation

 

at 100 °C

W

150

260

370

at 200 °C

W

300

520

750

Vacuum connection

vacuum connection measuring feedthrough needle valve for inert gas or air chamber leakage 

DN mm

16

16

16

DN mm

40

40

40

 mm

8

8

8

mbar.l.s–1

5.10-3

5.10-3

5.10-3

 

*) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.
**) Standard type is up to 70 °C, optional type is up to 99,9 °C.

Note: All technical data are related to 22 °C ambient temperature and ±10 % voltage swing (if not specified). Changes in the design and make reserved.