ISO9001:2015 & ISO14001:2015
LUMIstar OMEGA
Product Details
The LUMIstar® Omega is BMG LABTECH's powerful dedicated luminescence microplate reader, providing highly sensitive flash and glow assay readings. This DLReady-certified microplate reader can also come equipped with Simultaneous Dual Emission for BRET applications, and can be further upgraded to a multi-mode reader.
The high performance, dedicated luminescence microplate reader.
Details
The LUMIstar Omega is a dedicated reader for luminescence measurements, including BRET. The LUMIstar Omega provides highly sensitive readings of all luminescence applications in end-point and kinetic assays. Moreover, a built-in filter wheel and the optional Simultaneous Dual Emission detection make multiple wavelength detection and BRET applications possible.Features
- High performance microplate reader for luminescence
- Measurement at up to eight distinctive wavelengths
- Top and bottom reading
- BRET, optional with Simultaneous Dual Emission detection
- Up to two reagent injectors
- Upgradeable to FLUOstar Omega or POLARstar Omega including multiple detection modes and UV/Vis spectrometer
- Powerful MARS Data Analysis Software
Technical Specifications
Detection Modes |
Luminescence (flash and glow) BRET |
Measurement Modes |
Top and bottom reading Endpoint and kinetic measurements Sequential Multi Emission measurements Simultaneous Dual Emission measurements Ratiometric measurements |
Microplate Formats |
6 to 384-well plates, user-definable |
Microplate Carrier |
Robot compatible |
Detectors |
Low noise photomultiplier tube |
Wavelength Selection |
Optical Filters: Em wheels for 8 filters |
Optical Filters |
Emission filter wheel for 8 filters |
Optical Path Guides |
Top: Liquid filled light guides Bottom: Fiber optics |
Spectral Range |
Filters 240 - 740 nm for LUM |
Sensitivity* |
LUM < 20 amol/well ATP |
Read Times |
Flying mode (1 flash) 20 s (96), 55 s (384) |
Reagent Injection |
Up to 2 built-in reagent injectors Injection at measurement position (6 to 384-well) Individual injection volumes for each well (3 to 500 µL) Variable injection speed up to 420 µL / s Up to four injection events per well Reagent back flushing |
Shaking |
Linear, orbital, and double-orbital with user-definable time and speed |
Purge Gas Vent |
System to inject an atmosphere or to pull a vacuum into the reader |
Incubation |
+4 °C above ambient up to 45 °C or 65 °C The upper heating plate of the incubation chamber operates at 0.5 °C more than the lower plate. This prevents condensation build-up on the lid or sealer. |
Software |
Multi-user Reader Control and MARS Data Analysis Software included FDA 21 CFR Part 11 compliant |
Dimensions |
Width: 44 cm, depth: 48 cm, height: 30 cm; weight: 28 kg |
Optional Accessories |
|
Atmospheric Control Unit (ACU) |
Oxygen Regulation: 1 - 19% using nitrogen to purge. Carbon Dioxide Regulation: 0 - 20% using 100% moisture free carbon dioxide to purge. Dimensions (ACU only): Width: 44 cm, height: 17 cm and depth: 19 cm; Weight: 5 kg |
Stacker |
Magazines for up to 50 plates - continuous loading feature |
THERMOstar |
Microplate Incubator and Shaker |
Optical Filters |
Emission filter wheel for 8 filters |
* |
LOD = 3 x SD (20 blanks) / slope (6 pt std curve) Microplates: white for LUM 96 = 96-well microplates 384sv = 384-well small volume microplates 384g = 384-well glass bottom microplates |
Brochures
Write a review of LUMIstar OMEGA
Get in touch with us on 01939 234222
Need further assistance or can't find what you're looking for? Please complete all details of your enquiry and we will get back to you shortly.