ISO9001:2015 & ISO14001:2015
MMM VACUCELL EVO Vacuum Drying Oven
Product Details
NEW.....300°C for VACUCELL® EVO:
VACUCELL® EVO is the true leader in the field of high-end vacuum ovens. With it’s new option of temperature range up to 300°C it will help you to access high temperature vacuum applications to meet the strictest requirments.
The EVO Line:
- Intuitive control
- Microprocessor process control Fuzzy logic
- Multi-lingual communication
- Acoustic and visual alarm
- LED indicator of device functionality
- LCD display – 5,7 inches (14,5 cm)
- Graphic displaying of a new program
- Control through colour icons
- Touch display lock – protection from unauthorised access by a password
- Multi-level administration of users (corresponding to FDA 21 Part 11)
- Data coding and no-manipulability (according to FDA 21 Part 11)
- Up to 100 programs and up to 100 segments for each program
- Programming of temperature ramps, real time and cycling
- Annual data recording in graphic and numeric form
- Data export in online and offline mode
- Pre-set service programs for prompt diagnostics of failures
- Easy service diagnostics including remote access
- SD memory card, USB Host and interface RS 232 – included as a standard
- Connection: WiFi, USB Device or Ethernet interface with proper IP address for remote data transfer, control and diagnostics (optional equipment)
The VACUCELL® device is designed for perfect drying of materials, parts and samples up to constant weight, in vacuum, with possible protective atmosphere in inert gas. The VACUCELL® line is characterised by noise-free operation and ne heating and it offers equal and safe drying of thermo-labile, oxidation-sensitive or powder materials in laboratories, as well as shape-complicated products with many holes and threads in industry. The device offers high service comfort and exact regulation of temperature and vacuum and that is why it is suitable even for demanding and exact tests and processes. The device can be implemented in pharmaceutical, cosmetic, chemical, electro technical, petrochemical, air and tobacco industries, in the field of redevelopment technologies, space research and manufacture of health and medical means. On request, the device can be completed with a “Vacustation” base, with or without selected vacuum pump type. The devices of the VACUCELL® line meet the requirements of technical and legislative.
Technical Specifications
Inner space |
volume |
l |
22 |
55 |
111 |
width |
mm |
340 |
400 |
540 |
|
depth |
mm |
260 |
320 |
410 |
|
height |
mm |
300 |
430 |
480 |
|
External dimensions ((including door, handle, legs) |
width |
max. mm |
560 |
620 |
760 |
depth |
max. mm |
500 |
560 |
650 |
|
height |
max. mm |
780 |
910 |
960 |
|
Package – basic package only ECO |
width |
approx mm |
720 |
780 |
930 |
depth |
approx mm |
660 |
730 |
830 |
|
height (including palette) |
approx mm |
920 |
1050 |
1100 |
|
Package - case |
width |
approx mm |
720 |
780 |
930 |
depth |
approx mm |
660 |
730 |
830 |
|
height (including palette) |
approx mm |
960 |
1095 |
1150 |
|
Trays |
maximal number |
pc |
5 |
7 |
8 |
standard equipment |
pc |
2 |
2 |
2 |
|
minimal distance between trays/shelves |
mm |
36 |
43 |
43 |
|
usable area |
mm |
280×236 |
340×296 |
480×386 |
|
Maximal allowed loading of trays |
per 1 tray |
kg |
20 |
25 |
25 |
inside the device - in total |
kg |
35 |
45 |
65 |
|
Weight |
net |
approx kg |
65/68 |
98/101 |
130/133 |
brut (cartoon) |
approx kg |
76/91 |
111/186 |
145/218 |
|
Electric data – mains 50/60 Hz |
max. input |
kW |
0,8 |
1,2 |
1,8 |
stand by input |
W |
5/11 |
5/11 |
5/11 |
|
current for voltage *) |
A |
3,5 |
5,2 |
7,8 |
|
current for voltage *) |
A |
7 |
10,4 |
15,6 |
|
IP Code |
IP20 |
IP20 |
IP20 |
||
Temperature data |
|||||
Operation temperature |
from 5°C above ambient temperature |
to°C |
200/250 |
200/250 |
200/250 |
Temp. deviations acc. to DIN 12 880 from working temperature (Al racks, pressure 5-10 mbar) **) |
in space at 100°C |
± °C |
2 |
2 |
3 |
in space at 200°C |
± °C |
5 |
6 |
7 |
|
in time |
± °C |
0,4 |
0,4 |
0,4 |
|
Temp. deviations acc. to DIN 12 880 from working temperature (stainless racks, pressure 5-10 mbar) **) |
in space at 100°C |
± °C |
10 |
10 |
11 |
in space at 200°C |
± °C |
18 |
23 |
26 |
|
in time |
± °C |
0,5 |
1 |
1 |
|
Time of rise onto 98% voltage 230 V – Al racks, pressure 5-10 mbar |
onto temp. 100°C |
min |
60 |
65 |
110 |
onto temp. 200°C |
min |
80 |
85 |
130 |
|
Time of rise onto 98% voltage 230 V – stainless racks, press 5-10 mbar |
onto temp. 100°C |
min |
130 |
140 |
170 |
onto temp. 200°C |
min |
170 |
180 |
220 |
|
Heat emission |
at 100°C |
W |
150 |
260 |
370 |
at 200°C |
W |
300 |
520 |
750 |
|
Device noise level (without air pump) |
db |
0 |
0 |
0 |
|
Inert gas or air connection |
Needle valve ECO |
Ø mm |
8 |
8 |
8 |
Programmable filling EVO |
Ø mm |
8 |
8 |
8 |
|
Vacuum connection |
vacuum connection |
DN mm |
16 |
16 |
16 |
measuring feedthrough |
DN mm |
40 |
40 |
40 |
|
max. attainable vacuum |
mbar |
5.10-4 |
5.10-4 |
5.10-4 |
|
chamber untightness |
mbar.I.s-1 |
<5.10-3 |
<5.10-3 |
<5.10-3 |
Note: All technical data are related to 22°C ambient temperature and ± 10% voltage swing (if not specified). *) Not measured**) Heat transport to sample in shelves under vacuum is performed through shelves leads; that is why the above stated temperature variations apply to temperatures on shelves surfaces; the measuring sensors must be in perfect heat-conductive contact with the shelf surface. Samples placed on shelves must also be in perfect contact with shelves, the temperature of samples depends mainly on their physical characteristics and on contact with the shelf. The values may differ depending on specific charge and media parameters. Change in the design and make reserved.
Brochures
Write a review of MMM VACUCELL EVO Vacuum Drying Oven
Get in touch with us on 01939 234222
Need further assistance or can't find what you're looking for? Please complete all details of your enquiry and we will get back to you shortly.